Turbine Characterization - Siare Siaretron 4000 Service Manual

Icu ventilator, turbine-driven ventilation
Hide thumbs Also See for Siaretron 4000:
Table of Contents

Advertisement

6.2.1 Turbine Characterization

Turn the knob and select the function:
Turbine Characterization
1
Siaretron displays the activation
screen for turbine characterization:
2
Press START to begin
Press ESC to stop
The turbine characterization calibration is necessary in the following cases:
A significant deviation is noticed (ex. >10% or 2% end scale + 4% read value)
between the set pressure value and correspondent value effectively applied and
measured by ventilator.
In case of first calibration of ventilator or replacement of turbine.
Software upgrade or first start-up for important repairing (e.g.: replacement of turbine
or inspired flow sensor).
WARNING! Patient injury hazard
There are various procedures for device calibration.
Use the Encoder knob to select (turn) and activate (press) the desired
procedure.
If there is no CALIBRATION PROGRAMS confirm, the system
automatically switches to SELF TEST phase.