4.9
Ventilation phase
Press START soft key to begin the ventilation in the
selected mode with the most suitable PRP for the clinical
situation of the patient.
To returns in STAND-BY mode, please see 4.9.1.
In the upper section of the display you can find some values that will help you asses
the patient's clinical condition.
In the main section you can see the curves that show the trend of the respiratory
parameters.
On the right side of the screen there is a suitable lung icon that simulates the patient's
lungs, graphically displaying the respiratory cycle by alternatively switching the lungs
color.
4-36
Before starting the ventilation you have to:
carry out all ventilator hardware connections (medical gases, power
supply, patient circuit, ........)
carry out the preliminary checks (please see chapter 3.6)
set the language and the patient data
set and check the alarms limits
set the Physiological Respiratory Parameters and the operating mode
that match the patient's clinical situation best.
Based on the PRP set by the clinician according to the patient's
characteristics, the lung ventilator is able to display and measure a series
of values necessary for the patient's clinical evaluation.
User manual, DU3222102