Replacing The Lower Part Of The Housing (Wall/Tube Retaining Bracket) - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Compact controller measurement: inductive conductivity
Table of Contents

Advertisement

Replacing spare part units
15.2
Replacing the lower part
of the housing (wall/tube
retaining bracket)
Complete commissioning of
the controller
Once the lower part of the housing
has been replaced, it is necessary
to fully commission the measuring
and control point, as the new lower
part of the housing does not have
specific settings, only factory set‐
tings.
2
1
112
Fig. 46: Removing the wall/tube retaining
bracket
1.
Remove the wall/tube retaining
bracket. Pull the two snap-hooks (1)
outwards and push upwards
NOTICE!
Ribbon cable base
The base of the ribbon cable is
firmly soldered onto the PCB. The
base cannot be removed. Open
the base lock (3) to loosen the
ribbon cable, see Fig. 42
Fig. 47: Loosening the ribbon cable
2.
Undo four screws and open the
DULCOMETER
A0273
troller
3.
Open the right and left lock (3)
(arrows) on the base and pull the
ribbon cable (1) out of the base.
The catches (2 and 4) are used to
aligned the two halves of the
housing.
2.
1.
4.
Compact Con‐
®
3.