Calibrating The Zero Point - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Compact controller measurement: inductive conductivity
Table of Contents

Advertisement

Operating menus
9.1.3

Calibrating the zero point

Dry the sensor and keep it free from electromagnetic fields
If the zero point of the sensor is to be calibrated, it is essential that the sensor is
removed, rinsed and dried before calibration. No electromagnetic fields may influ‐
ence the sensor as these electromagnetic fields could distort calibration. Keep
devices that transmit radiation away from the sensor, such as mobile phones,
WLAN routers, sources of high voltage, transformers etc.
CAL ZERO
ZERO = 710.793 µS
CAL=ACCEPT OK=RESET
Fig. 26: Calibrating the zero point / The [CAL] [ZERO] default value corresponds to the
default value of the sensor selected.
Tab. 9: Zero point calibration limits
Sensor
ICT 1
ICT 2
ICT 5
CLS52
Manual
64
CAL
CELLCONST
TCOEFF
ZERO
CAL ZERO
Put LF-Sensor
in air
CAL=ACCEPT
CAL ZERO
SENSOR= 762.µS
STABLE (10 sec)?
CAL=ACCEPT
Value
65 µS
4 µS
30 µS
4 µS
50 uS
A1942