Replacing The Lower Part Of The Housing (Control Panel Installation) - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Compact controller measurement: inductive conductivity
Table of Contents

Advertisement

15.3
Replacing the lower part
of the housing (control
panel installation)
Complete commissioning of
the controller
Once the lower part of the housing
has been replaced, it is necessary
to fully commission the measuring
and control point, as the new lower
part of the housing does not have
specific settings, only factory set‐
tings.
NOTICE!
Ribbon cable base
The base of the ribbon cable is
firmly soldered onto the PCB. The
base cannot be removed. Open
the base lock (3) to loosen the
ribbon cable, see Fig. 42
1.
A0352
Replacing spare part units
Fig. 52: Loosen the ribbon cable from the
base
1.
Undo four screws and open the
DULCOMETER
troller
2.
Open the right and left lock on the
base and pull the ribbon cable (1)
out of the base.
2.
Fig. 53: Dismantling the hinge
3.
Remove the screw (2), unclip the
hinge (1) on the lower part of the
controller housing (arrows) and
remove the hinge
1.
Fig. 54: Removing the strain relief
4.
Remove the strain relief (2).
Remove the screws (1) to do so.
®
Compact Con‐
1.
2.
1.
115