4.
Press
and use the
firm with
.
5.
Enter the temperature coefficient of the calibration solution.
The temperature coefficient of the calibration solution is specified on the
storage tank for the calibration solution.
Confirm with
6.
Now dip the sensor into the calibration solution and gently move the sensor.
7.
Wait until the conductivity and temperature measured value has stabilised.
Press
.
ð The conductivity value measured is displayed.
8.
Now use the
ance with the conductivity value specified on the calibration solution.
ð This first calibration in a known calibration solution delivers a cell constant of
[CC]
[Test1]
Calculating the valid value
9.
Now use the following formula to calculate the valid value for
[MATCH_CC] = [MATCH_CC]
10.
Enter this calculated value for
Recalibrate using a calibration solution that corresponds approximately to the conduc‐
tivity of your application
11.
Select the calibration menu and select
12.
Press
and use the
firm with
.
13.
Enter the temperature coefficient of the calibration solution.
The temperature coefficient of the calibration solution is specified on the
storage tank for the calibration solution.
or
key to move the cursor to
.
,
or
keys to enter the conductivity value measured, in accord‐
. The setting was as follows:
[Test1]
[MATCH_CC] in the [INPUT] menu.
or
key to move the cursor to
[MATCH_CC]
[CC]
[CC]
*
/
[Test1]
Manufacturer's figure
[CELLCONST] .
Operating menus
[CELLCONST] and con‐
100.0
[Test1]
[MATCH_CC] :
[CELLCONST] and con‐
79