Cable Cross-Sections And Cable End Sleeves; Electrical Connection Of The Conductivity Sensor - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Compact controller measurement: inductive conductivity
Table of Contents

Advertisement

5.4.1

Cable Cross-Sections and Cable End Sleeves

Without cable end
sleeve
Cable end sleeve
without insulation
Cable end sleeve
with insulation
5.4.2
Electrical connection of the
conductivity sensor
CAUTION!
Length of the sensor cable
The sensor is supplied with a fixed
cable or a measuring line.
Possible consequence: Slight or
minor injuries. Material damage.
Set the length of the sensor cable
[INPUT] menu so that the Pt
in the
100 sensor measures the tempera‐
ture correctly.
Ensure that all conductivity sen‐
sors connected to the controller
have a shielded sensor cable.
Minimum cross-sec‐
tion
0.25 mm
2
0.20 mm
2
0.20 mm
2
Assembly and installation
Maximum cross-
section
1.5 mm
2
1.0 mm
2
1.0 mm
2
Stripped insulation
length
8 - 9 mm
10 - 11 mm
31