ProMinent DULCOMETER Assembly And Operating Instructions Manual page 114

Compact controller measurement: inductive conductivity
Table of Contents

Advertisement

Replacing spare part units
Fig. 50: Fix the ribbon cable
14.
Push and lock the ribbon cable (1)
in its base. The catches (2 and 4)
are used to aligned the two halves
of the housing.
15.
Screw the top part of the controller
housing onto the lower part of the
DULCOMETER
troller housing
16.
Re-check that the seal is seated
properly. IP 67 degree of protection
(wall/pipe-mounting) can only be
provided if the installation is correct
1
3
114
2.
1.
4.
®
Compact Con‐
2
Fig. 51: Suspend and secure the
3.
DULCOMETER
17.
Suspend the DULCOMETER
Compact Controller at the top (1) in
the wall/tube retaining bracket and
push using light pressure at the
bottom (2) against the wall/pipe
retaining bracket. Then press
upwards (3) until the
DULCOMETER
troller audibly snaps into position
A0275
Compact Controller
®
Compact Con‐
®
®

Hide quick links: