ProMinent DULCOMETER Assembly And Operating Instructions Manual page 117

Compact controller measurement: inductive conductivity
Table of Contents

Advertisement

15.
Use pliers to break off the catches.
They are not needed for control
panel installation
Position the profile seal evenly
around the top edge of the lower
part of the DULCOMETER
pact Controller housing. Arrange
the flaps (1) as shown in the figure
ð Ensure that the profile seal
evenly surrounds the top edge
of the housing.
16.
Insert the lower part of the
DULCOMETER
troller housing with the profile seal
from behind into the cut-out and
use three screws to secure it in
place
1.
®
Com‐
Compact Con‐
®
2.
1.
Replacing spare part units
Fig. 57: Fit the profile seal onto the top
part of the controller housing
17.
Position the profile seal (arrow)
evenly into the groove in the top
part of the DULCOMETER
pact Controller housing. Arrange
the flaps (3) as shown in the figure
18.
Secure the strain relief (2) using
two screws (1)
19.
Fit the hinge
Fig. 58: Pushing and locking the ribbon
cable in its base
20.
Push and lock the ribbon cable (1)
in its base
21.
Screw the top part of the controller
housing onto the lower part of the
DULCOMETER
A0351
troller housing
22.
Re-check that the profile seals are
fitted properly
ð IP 54 degree of protection can
only be provided if the control
panel is mounted correctly
®
Com‐
1.
®
Compact Con‐
117
A0352