ProMinent DULCOMETER Assembly And Operating Instructions Manual page 107

Compact controller measurement: inductive conductivity
Table of Contents

Advertisement

The mains connection is isolated from other switching parts by reinforced insulation. The
unit has no mains switch; a fuse is fitted.
Output relay (P-relay)
Load capacity of switching contacts
Outputs galvanically isolated from other switching parts by reinforced insulation.
Digital input
Open circuit voltage
Short circuit current
Max.switching frequency
NOTICE!
Do not supply with voltage
To connect an external semiconductor or mechanical switch.
mA output
0 ... 20 mA
Current range
0 ... 20.5 mA
In the event of
0 and/or 23 mA
a fault
Max. load
480 W at 20.5 mA
Max. output
19 V DC
voltage
Overvoltage-
±30 V
resistant up to
Output accu‐
0.2 mA
racy
5 A; no inductive loads
22 V DC max.
6.5 mA
Static. For switching processes like
'PAUSE' , 'HOLD' etc.
4 ... 20 mA
3.8 ... 20.5 mA
3.6 and/or 23 mA
Electrical data
manual
0 ... 25 mA
107