ProMinent DULCOMETER Assembly And Operating Instructions Manual page 62

Compact controller measurement: inductive conductivity
Table of Contents

Advertisement

Operating menus
CAL TEMPCOEFF
TCOEFF=0.00%/°C
CAL=OLD
Fig. 25: Calibration of the temperature coefficient
I.
If the temperature change is greater
than 2°C, the message changes to
[WAIT]
II.
If the temperature is within the speci‐
fied range, the message changes to
[STABLE ?]
III. When a stable final temperature has
been reached, the message changes
[ACCEPT?] . Calibration can now be
to
terminated manually.
62
CAL
CELLCONST
TCOEFF
CAL TEMPCOEFF
CAL_T1= 22.0 °C
CAL=CONTINUE
CAL TEMPCOEFF
CHANGE TEMP
T2 > 32.0 °C
or
T2 < 12.0 °C
I.
CAL TEMPCOEFF
WAIT
T2 > 32.0 °C
or
T2 < 12.0 °C
II.
CAL TEMPCOEFF
STABLE ?
CAL_T2= 43.0 °C
CAL=CONTINUE
III.
CAL TEMPCOEFF
ACCEPT ?
CAL_T2= 43.0 °C
CAL=CONTINUE