ProMinent DULCOMETER Assembly And Operating Instructions Manual page 67

Compact controller measurement: inductive conductivity
Table of Contents

Advertisement

Setting
Display
Starting
value
[LIMIT ↑]
0.02 S/cm
[LIMIT ↓]
0.01mS/cm 0,001
[HYST. ]
5 %
[TLIMIT ↑]
30.0 °C
°C
[TLIMIT ↓]
10.0 °C
°C
[TLIMIT ↑]
86.0 °F
°F
[TLIMIT ↓]
32.0 °F
°F
[TIMELIM.]
0 min =
OFF
[TDS] or [SAL] is set in the continuous
If
display, the setting values for
[TLIMIT↓] are hidden in the [LIMIT]
and
menu:
[TLIMIT↓] can be changed if the con‐
n
tinuous display is showing
[RES] .
[TLIMIT↑] is fixed at 40 °C (with TDS)
n
and 35 °C (with SAL). If the value set
[Cond_I] for [TLIMIT↑] is less than
at
this value, this setting is retained.
Possible values
Increment
Lower value Upper value Remark
0,001
0.000
uS/cm
0.000
uS/cm
1 %
1 %
0.1 °C
0.0 °C
0.1 °C
0.0 °C
0.1 °F
32.0 °F
0.1 °F
32.0 °F
1 min
0
Hysteresis: the hysteresis is specified as a
[TLIMIT↑]
%, as an absolute indication is impossible
due to the extent of the measuring range.
The indication refers to the values given
under
[Cond_I] or
Operating menus
2.000 S/cm
Upper limit value
2.000 S/cm
Lower limit value
20 %
Hysteresis of
limit values
150.0 °C
Upper limit value
correction value
°C
150.0 °C
Lower limit value
correction value
°C
302.0 °F
Upper limit value
correction value
°F
302.0 °F
Lower limit value
correction vari‐
able °F
999
Checkout time
after occurrence
of a limit value
transgression
[LIMIT↑] and [LIMIT↓] .
67