Calibration Of The Temperature Coefficient - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Compact controller measurement: inductive conductivity
Table of Contents

Advertisement

Tab. 8: Sensor status
Display
Meaning
[OK]
In order
[WRN]
Warning
[ERR]
Error
9.1.2

Calibration of the temperature coefficient

Conductivity sensors with temperature element
You can only calibrate the temperature coefficient with conductivity sensors with a
temperature element, because it is impossible to calculate the temperature coeffi‐
cient without measuring the temperature.
Temperature change
It is recommended that the temperature is changed by no more than 0.5 °C per
minute, or with a temperature change of e.g. 10 °C you will need to wait for a min‐
imum of 20 minutes before calibration.
Status
Cell constant = 0.005 / 99.9
none
Cell constant < 0.005 or cell constant > 100
Operating menus
61