Operating menus
Confirm with
14.
Now dip the sensor into the calibration solution and gently move the sensor.
15.
Wait until the conductivity and temperature measured value has stabilised.
Press
.
ð The conductivity value measured is displayed.
16.
Now use the
ance with the conductivity value specified on the calibration solution.
ð The cell constant
the controller's memory.
80
.
,
or
keys to enter the conductivity value measured, in accord‐
[CC] has now been correctly ascertained and is available in