ProMinent DULCOMETER Assembly And Operating Instructions Manual page 99

Compact controller measurement: inductive conductivity
Table of Contents

Advertisement

Tab. 10: Error messages
[Error ]
Error mes‐
sage
[Warning]
[RANGE↓]
[E]
[RANGE↑]
[E]
[T
[E]
RANGE↓]
[T
[E]
RANGE↑]
[CAL
[E]
ERROR]
[CHECK‐
[E]
TIME]
[mA
[E]
RANGE↑]
[mA
[E]
RANGE↓]
[LIMIT↑]
[W]
[LIMIT↓]
[W]
[T LIMIT↑]
[W]
[T LIMIT↓]
[W]
[LIMIT
[E]
ERR]
[TLIMI‐
[E]
TERR]
[NO CAL]
[W]
[CON‐
[E]
TACTIN]
[TDS↑]
[W]
[SAL↑]
[W]
Brief description of fault
The measured variable falls below the measuring range
The main measured variable exceeds the measuring range
Temperature value falls below the measuring range
Temperature value exceeds the measuring range
Calibration error during last calibration performed by the user
Ä Chapter 10.7 'Checkout time control'
See chapter
on page 93
The current emitted at the mA standard signal output is
higher than 20 mA. This does not apply with the issue of the
23 mA error current
The current emitted at the mA standard signal output is below
0/4 mA. This does not apply with the issue of the 0/3.6 mA
error current
Measured variable is higher than the limit set
Measured variable is lower than the limit set
Correction measured variable higher than the limit set
Correction measured variable higher than the limit set
Ä Chapter 10.6 'Checkout time for measured
See chapter
variable and correction variable' on page 93
Ä Chapter 10.6 'Checkout time for measured
See chapter
variable and correction variable' on page 93
No calibration has been performed by the user
Alarm triggered by the contact input (
[ALARM=on] selected)
[TDS] value is too high. Continuous display: > 2000
The
[SAL] value is too high. Continuous display: > 70
The
Maintenance
[INPUT] menu:
99