ProMinent DULCOMETER Assembly And Operating Instructions Manual page 59

Compact controller measurement: inductive conductivity
Table of Contents

Advertisement

Calibration compared to a calibration solution
1.
Press
and use the
firm with
.
2.
Enter the temperature coefficient of the calibration solution.
The temperature coefficient of the calibration solution is specified on the
storage tank for the calibration solution.
Confirm with
3.
Now dip the sensor into the calibration solution and gently move the sensor.
4.
Wait until the conductivity and temperature measured value has stabilised.
Press
.
ð The conductivity value measured is displayed.
5.
Now use the
ance with the conductivity value specified on the calibration solution.
ð If calibration has been completed successfully, the controller stores the deter‐
mined values for the cell constant and the error checks relating to the meas‐
ured value are restarted. The numerical setting range of the cell constant is
not limited.
6.
Press
twice to return to the continuous display.
or
key to move the cursor to
.
,
or
keys to set the conductivity value measured, in accord‐
Operating menus
[CELLCONST] and con‐
59