The inductive conductivity measured at
the fluid temperature is converted to the
reference temperature
Changing the reference tem‐
perature
The temperature coefficient
[TCOEFF] has to be recalibrated if
the reference temperature [TREF]
is changed.
Adjustable process for temperature com‐
pensation
[off]
n
Temperature compensation is
switched off. It is measured
based on the set reference tem‐
perature.
[lin]
n
Linear temperature compensa‐
Ä Chapter 10.5 'Tem‐
tion, see
perature correction variable'
on page 92 , for the temperature
range permitted for the sensors.
The reference temperature
[TREF] can be set between 15 °C
and 30 °C.
[nLF]
n
Non-linear temperature compen‐
sation according to DIN EN
27888 for natural water, between
0 °C ... 35 °C. The reference tem‐
[TREF] can be switched,
perature
20 °C / 25 °C.
[TREF] .