Temperature
Thermocouple J
Range: -200 to 1,000 ˚C
Resolution: 0.1 ˚C
Accuracy:
-200 to – 60 ˚C ±0.5 %
60 to 1,000 ˚C ±0.5 %
-60 to 60 ˚C ±0.5 ˚C
Cold junction
compensation: ±0.3 ˚C
Temperature PT-100
(2-wire)
Range: -200 to 400 ˚C
Resolution: 0.1 ˚C)
Accuracy:
-200 to - 60 ˚C ±0.5 %
60 to 400 ˚C ±0.5 %
-60 to 60 ˚C ±0.3 ˚C
4 to 20 mA
Range: 4 to 20 mA
Resolution: 4.76 μA
Accuracy: ±0.5 %
Loop impedance: 21 Ω
Maximum load: 30 mA,
5.2 V
Temperature
Thermocouple K
Range: -200 to 1,000 ˚C
Resolution: 0.1 ˚C
Accuracy:
-200 to – 60 ˚C ±0.5 %
60 to 1,000 ˚C ±0.5 %
-60 to 60 ˚C ±0.5 ˚C
Cold junction
compensation: ±0.3 ˚C
Pulse Counter
(Input 4 only)
Range: 1 to 65,536
counts
Resolution: 1 count
Frequency range: 0 to 4
KHz
Input signal: 0 to 5 V
Input impedance: 470 Ω
0 to 1 V
Range: 0 to 1 V
Resolution: 200 μV
Accuracy: ±0.5 %
Input impedance: 25
MΩ
Maximum voltage: 5.2 V
185
4% (Logger LCD)
Temperature
Thermocouple T
Range: -200 to 400 ˚C
Resolution: 0.1 ˚C
Accuracy:
-200 to – 60 ˚C ±0.5 %
60 to 400 ˚C ±0.5 %
-60 to 60 ˚C ±0.5 ˚C
Cold junction
compensation: ±0.3 ˚C
Frequency
(Input 4 only)
Range: 20 Hz to 4 KHz
Input signal: 0 to 5 V
Input impedance: 470 Ω
0 to 50 mV
Range: 0 to 50 mV
Resolution: 3 μV
Accuracy: ±0.5 %
Input impedance: 25
MΩ
Maximum voltage: 5.2 V