Introduction To Datanet Calibration - Fourier DataNet User Manual

Wireless intelligent logging network
Table of Contents

Advertisement

The Calibration option menu options are:
Calibration: Opens the main Calibration dialog
Save Calibration: Saves the logger's current calibration
settings
Load Calibration: Loads a logger's previously saved
calibration settings.
Reset Calibration: Resets the logger's calibration
settings so that the raw hardware data is obtained,
without any software calibration applied to these values.
Restore Factory Calibration Default: Restores the
logger's calibration settings to the factory calibration i.e.
the calibration settings it received prior to shipment from
Fourier.
4.14.1. Introduction to DataNet
Calibration
Although the DataNet data loggers come fully calibrated, the
software enables you to calibrate any of the DataNet sensors,
on any input. The calibration parameters are sent to the data
logger and stored in its memory.
DataNet employs two different calibration methods: Two-
point calibration and Offset calibration.
Each sensor can be calibrated using the Two-point calibration
method, and then tweaked using offset calibration, except for
the Thermocouples. These sensors require offset calibration
only.