Thermocouples you will only have the option to perform an
Offset calibration.
Enter the offset value in the text box and click Send
Calibration. Continue to adjust the offset accordingly until
satisfied with the readings.
4.14.5. Setting the Offset to a Specific
Input
After setting the offset to all inputs simultaneously, you can
set the offset of each input separately for even better
accuracy.
To set the offset of a specific input:
1. Measure two known reference values making sure to use
the input you need to calibrate.
2. Enter the offset value as described in section 4.14.4
above, making sure to select the correct input number
instead of All.
3. You can repeat the procedure for every input.
4.14.6. Calibrating the Internal
Temperature Sensor on the
DL910
The DNL910's internal temperature sensor is a PT-100
sensor. Calibration of this sensor is performed by calibrating
the external PT-100 input i.e. two-point calibration followed by
offset calibration.
1. In the Calibration dialog, select PT-100 in the Sensor
drop-down menu and select All inputs.
2. Press Setup to proceed.