Calibrating The Data Logger; Figure 51: Calibration Dialog - Fourier DataNet User Manual

Wireless intelligent logging network
Table of Contents

Advertisement

4.14.2.

Calibrating the Data Logger

Figure 51: Calibration dialog

1. From the Map View, open the Logger context menu and
select Stop.
2. Select Calibration > Calibrate and enter the password in
the Password dialog.
3. Select the sensor you wish to calibrate from the Sensor
drop-down menu.
4. If calibrating an external sensor, select the inputs to
calibrate. Select All or an individual input, 1 to 4.
5. Click Setup to send the sensor setup to the logger.
Note: The calibration process does not delete the logger's
setup prior to calibration. Once Calibration is complete the
original logger setup will be restored.