Aoutput Group - YOKOGAWA WT200 User Manual

Digital power meter
Table of Contents

Advertisement

14.3 Commands

14.3.2 AOUTput Group

The commands in the AOUTput group are used to make settings relating to, and inquires about D/A output. This
allows you to make the same settings and inquiries as can be set using the lower menus of [OUTPUT]-"dA" or
[INTEG SET]-"dAtimE". The commands in this group are available only if the /D4 option is installed.
:AOUTput
:
AOUTput?
Function
Queries all the settings relating to D/A output.
Syntax
AOUTput?
Example
AOUTPUT?→:AOUTPUT:CHANNEL1 V,1;
CHANNEL2 A,1;CHANNEL3 W,1;
CHANNEL4 VHZ,1;IRTIME 1,0,0
AOUTput:CHANnel<x>
Function
Sets the D/A output item, or queries the current
setting.
Syntax
AOUTput:CHANnel<x> {<normal measurement
function>,(<NRf>|ELEMent<1>)|OFF}
<x>=1 to 4 (for /DA4)
<normal measurement function>={V|A|W|VA|
VAR|PF|DEGRee|VHZ|AHZ|WH|WHP|
WHM|AH|AHP|AHM|MATH|VPK|APK}
Example
AOUTPUT:CHANNEL1 V,1
AOUTPUT:CHANNEL1?→:AOUTPUT:CHANNEL1V,1
AOUTPUT:CHANNEL2?→:AOUTPUT:CHANNEL2 OFF
14-14
;
PRESet
<Space>
NORMal
INTEGrate
CHANnel
<x>
<Space>
IRTime
<Space>
<Character string>
?
?
<Normal meas. function>
OFF
?
<NRf>
,
<NRf>
AOUTput:IRTime
Function
Syntax
Example
AOUTput:PRESet
Function
Syntax
Example
Description
,
<NRf>
ELEMent
<x>
,
<NRf>
Sets the preset integration time for D/A output
of integrated values, or queries the current
setting.
AOUTput:IRTime {<NRf>,<NRf>,<NRf>|
<String>}
{<NRf>,<NRf>,<NRf>}=0,0,0 to 10000,0,0
{<String>}=HHHHH:MM:SS
HHHHH hours
MM minutes
SS seconds
AOUTPUT:IRTIME 1,0,0
AOUTPUT:IRTIME "1:00:00"
AOUTPUT:IRTIME?→:AOUTPUT:IRTIME 1,0,0
Initializes the output items for D/A output.
AOUTput:PREset {NORMal|INTEGrate}
NORMal=default for normal measurement
INTEGrate=default for integration
AOUTPUT:PRESET NORMAL
See section 9.3 for a description of default D/A
output items for normal measurement and
integration.
IM 253421-01E

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents