Instrument Sanitize - Keysight N6705C Operating And Service Manual

Dc power analyzer
Table of Contents

Advertisement

Instrument Sanitize

This procedure is not recommended for use in routine applications because of the
possibility of unintended loss of data.
Sanitize removes all user-data from the instrument and restores the factory-shipped settings. Refer to
Non-Volatile Settings
for a list of the factory settings.
This procedure is typically used to prepare the instrument for removal from a secure area. It writes all
zeros to flash memory and then performs a full chip erase as per the manufacturer's data sheet.
Identification data such as instrument firmware, model number, serial number, MAC address, and
calibration data is not erased. After the data is cleared, the instrument is rebooted.
From the Front Panel
Log into the Administrative Tools menu. If you cannot access the Admin menu, it may be password
protected. Then select Sanitize and press the Sanitize button.
From the Remote Interface
SYST:SEC:IMM
Turn-on After Sanitization
The first time the unit is turned on after it has been sanitized, several NVRAM checksum errors will be
generated. These errors annunciate the fact that two files were missing, which have been recreated
with default values. The next time the unit tuns on, there should be no errors.
Keysight N6705C Operating and Service Guide
9 Service and Maintenance
439