Parameter
–524,287 to 2,000,000,000|MIN|MAX, *RST 0
Specifies -2048 offset points: SENS:SWE:OFFS:POIN -2048, (@1)
SENSe:SWEep:POINts <value>, (@<chanlist>)
SENSe:SWEep:POINts? [MIN|MAX,] (@<chanlist>)
Defines the number of points in a measurement. The number of points depends on the line frequency
(50 Hz or 60 Hz). The default number of points result in a measurement of 1 NPLC.
Parameter
1 to 524,288|MIN|MAX,
*RST 3255 (60 Hz); 3906 (50 Hz)
Specifies 2048 points: SENS:SWE:POIN 2048, (@1)
SENSe:SWEep:TINTerval <value>, (@<chanlist>)
SENSe:SWEep:TINTerval? [MIN|MAX,] (@<chanlist>)
Defines the time period between measurement samples.  Units are in seconds. Values are rounded to
the nearest 20.48 microsecond increment. Below 20.48 microseconds, values are rounded to the
nearest 10.24 or 5.12 microsecond increment respectively.
0.00000512 to 40,000|MIN|MAX, *RST 0.00000512
Specifies an interval of 1 ms between points: SENS:SWE:TINT 0.001, (@1)
SENSe:SWEep:TINTerval:RESolution RES20|RES40
SENSe:SWEep:TINTerval:RESolution?
Sets the resolution used in setting the time interval between digitized measurements. Applies for time
intervals above 20.48 microseconds. External data log measurements also use this setting.
RES20 -- rounds all values above 20.48 microseconds to the nearest 20.48 µs multiple.
RES40 -- rounds all values above 20.48 microseconds to the nearest 40.96 µs multiple.
Parameter
RES20|RES40, *RST 20.48 µs
Specifies a 40 µs resolution: SENS:SWE TINT:RES RES40
Keysight N6705C Operating and Service Guide
Parameter
7 SCPI Programming Reference
Typical Return
<offset points>
Typical Return
<points>
Typical Return
<time interval>
Typical Return
RES20 or RES40