Keysight N6705C Operating And Service Manual page 268

Dc power analyzer
Table of Contents

Advertisement

7 SCPI Programming Reference
Parameter
The maximum current of the range.
Calibrates the positive current limit: CAL:CURR:LIM:POS 2, (@1)
CALibrate:CURRent:MEASure <value>, (@<channel>)
Calibrates the current measurement. The value entered selects the range that is being calibrated.
Parameter
The maximum current of the range.
Calibrates the current measurement range: CAL:CURR:MEAS 10.2, (@1)
CALibrate:CURRent:PEAK (@<channel>)
Calibrates the peak current limit. 
Parameter
(none)
Calibrates the peak current limit: CAL:CURR:PEAK (@1)
CALibrate:DATA <value>
Enters the calibration value read by the external meter. You must first select a calibration level for the
value being entered. Data values are expressed in base units - either volts or amperes, depending on
which function is being calibrated.
Parameter
The measured voltage or current value.
Specifies the measured value 10.01: CAL:DATA 10.01
CALibrate:DATE <"date">, (@<channel>)
CALibrate:DATE?
Stores the calibration date in nonvolatile memory. Enter any ASCII string up to 16 characters. The
query returns the date. If no date is stored, an empty quoted string ("") is returned.
Parameter
<"date">String program data. Enclose string parameters in
single or double quotes.
Enters the calibration date: CAL:DATE "12/12/12"
The firmware does not interpret the string format. The command only stores the calibration date.
l
268
Typical Return
(none)
Typical Return
(none)
Typical Return
(none)
Typical Return
(none)
Typical Return
Date of last calibration
Keysight N6705C Operating and Service Guide