Parameter
FIXed|STEP|LIST|ARB, *RST FIXed
Sets the voltage mode to Step: VOLT:MODE STEP, (@1)
[SOURce:]VOLTage:PROTection[:LEVel] <value>, (@<chanlist>)
[SOURce:]VOLTage:PROTection[:LEVel]? [MIN|MAX,] (@<chanlist>)
This command does NOT apply to models N678xA SMU. Refer to
[SOURce;]VOLTage:PROTection:REMote and
[SOURce:]VOLTage:PROTection:REMote:NEGative.
Sets the over-voltage protection level.Units are in volts.
If the output voltage exceeds the OVP level, the output is disabled and the Questionable Condition
status register OV bit is set.
Parameter
0 – maximum|MIN|MAX, *RST MAX
The maximum value is 120% of the voltage rating of
the power module.
Sets the over-voltage protection to 60 V: VOLT:PROT 60, (@1)
An over-voltage condition can be cleared with the OUTput:PROTection:CLEar command after the
l
cause of the condition has been removed.
The query returns the programmed voltage protection level in the form +n.nnnnnnE+nn for each
l
channel specified. Multiple responses are separated by commas.
[SOURce:]VOLTage:PROTection:DELay[:TIME] <value>, (@<chanlist>)
[SOURce:]VOLTage:PROTection:DELay[:TIME]? [MIN|MAX,] (@<chanlist>)
Sets the over-voltage protection delay. The over-voltage protection function will not be triggered
during the delay time. After the delay time has expired, the over-voltage protection function will be
active. This prevents momentary changes in output status from triggering the over-voltage protection
function. Programmed values can range from 60 microseconds to 5 milliseconds with a resolution of
20.48 microseconds.
Parameter
60 μs - 5 ms | MIN | MAX, *RST 0
Sets the protection delay to 1 millisecond: VOLT:PROT:DEL 0.001, (@1)
Keysight N6705C Operating and Service Guide
7 SCPI Programming Reference
Typical Return
FIX, STEP, LIST, or ARB
Typical Return
<over-voltage level>
Typical Return
<delay value>