7 SCPI Programming Reference
Calibration mode is locked out by the calibration switch.
104 Bad sequence of calibration commands
Calibration commands have not been entered in the proper sequence.
105 Unexpected output current
The measured output current is outside the acceptable range.
106 Zero measurement out of range error
The "zero" measurement value is outside the acceptable range.
107 Programming cal constants out of range
The programmed calibration constant is outside the acceptable range.
108 Measurement cal constants out of range
The measurement calibration constant is outside the acceptable range.
109 Over voltage cal constants out of range
The over voltage calibration constant is outside the acceptable range.
110 Wrong V+I
The instrument was unable to set the correct voltage or current value.
111 Aux vloc cal constants out of range
Calibration constants on the internal auxiliary local ADC are outside the acceptable range.
112 Aux vrem cal constants out of range
Calibration constants on the internal auxiliary remote ADC are outside the acceptable range.
113 Aux imon cal constants out of range
Calibration constants on the internal auxiliary imon ADC are outside the acceptable range.
114 Wrong status
An incorrect status function has been reported.
115 Output track resistance out of range
The output track resistance is out of range. Re-calibrate the output module.
116 Locked out by internal switch setting
This function has been locked out by an internal switch.
117 Calibration error
A calibration error has occurred. Do not save calibration constants. Try re-calibrating the unit.
200 Hardware error channel <1>
Keysight N6705C Operating and Service Guide
361
Need help?
Do you have a question about the N6705C and is the answer not in the manual?