Chapter 6—Alarms and Troubleshooting
Message
Purpose of Message
Base Na+ greater than
Entered Base Na+ higher is than
max. value
allowed.
Base Na+ has been set
The operator has attempted to set a
to min.
Base Na+ lower than allowed.
Bic Pump Alarm
This is a pump failure warning.
Bic Pump Always EOS
This is a pump failure warning.
Bic Pump No EOS
This is a pump failure warning.
Bic Conn Out of Port
The blue bicarbonate wand/connector
is out of its port.
Bicarbonate greater than
The operator has attempted to set a
max. value
Bicarbonate level higher than allowed.
Bicarbonate has been set
The operator has attempted to set a
to min.
Bicarbonate level lower than allowed.
Blood flow unstable
When attempting to start an OLC test,
certain conditions are necessary,
including stable blood flow rate.
Blood Leak not Calib
The blood leak detector is not in
calibration.
149
Type
Action Required
Dialog
The Base Na+ will be set to the highest allowed Na+ level. Press
Message
CONFIRM to clear message and accept the maximum allowed value.
Verify that the value is acceptable or enter new value.
Dialog
The Base Na+ will be set to the lowest allowed Na+ level. Confirm if
Message
value is acceptable or enter new value.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly, turn machine power off and back on. If alarm is not
cleared, return the blood to the patient if alarm occurs during
treatment. Take the machine out of service.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly, turn machine power off and back on. If alarm is not
cleared, return the blood to the patient if alarm occurs during
treatment. Take the machine out of service.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly, turn machine power off and back on. If alarm is not
cleared, return the blood to the patient if alarm occurs during
treatment.Take the machine out of service.
Warning
Connect blue (bicarbonate) wand/connectors into the blue rinse port.
Verify the concentrate selection.
Dialog
The Bicarbonate will be set to the highest allowed bicarbonate level.
Message
Press CONFIRM to clear message and accept the maximum allowed
value. Verify that the value is acceptable or enter new value.
Dialog
The Bicarbonate will be set to the lowest allowed bicarbonate level.
Message
Confirm if value is acceptable or enter new value.
Advisory
Wait a minute or so and start the OLC test again.
Alarm
Return the blood to the patient if alarm occurs during treatment. Take
the machine out of service and alert a qualified service technician.
2008K Hemodialysis Machine Operator's Manual Rev P