Fresenius Medical Care 2008K Operator's Manual page 185

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Chapter 6—Alarms and Troubleshooting
Message
Purpose of Message
Verifying temp sensors
The machine will go into bypass for
about 8 minutes while the
temperature sensors are verified.
RTD will pause.
Wait, OLC Aborting
Changed blood pump flow rate,
changed dialysate flow rate, or
unstable conductivity caused on line
clearance test to stop.
Wait: Rinsing Line
The machine is rinsing the
concentrate lines prior to a cleaning
or disinfecting program.
WD: 24v Rcvr Err Long
Electronic self-test failure.
WD: 24v Rcvr Err Short
Electronic self-test failure.
WD: Fail Long Pulse
Electronic self-test failure.
WD: Fail Short Pulse
Electronic self-test failure.
185
Type
Action Required
Advisory
Advisory only. No action is required
Advisory
Wait until stable conditions for OLC test to begin again
Advisory
Advisory only. No action is required. Line rinsing takes about 45
seconds.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
2008K Hemodialysis Machine Operator's Manual Rev P