Fresenius Medical Care 2008K Operator's Manual page 86

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Chapter 3—Setting Treatment Parameters
Table 19 – Control Panel Heparin on/off Key
Key
The Heparin Delivery System
Warning! The correct syringe type must be selected to ensure an accurate infusion.
To prepare the heparin delivery system using the features on the "Heparin" screen:
1) Fill the
syringe selected
from the menu
with the
prescribed
amount of
heparin for the
entire treatment.
2) In the "Heparin"
screen, touch the
Syringe button.
The syringe menu
opens.
3) Use the Up/Down
keys on the data
entry keypad to
scroll the drop-down
menu until the
correct syringe type
is highlighted.
4) Press CONFIRM.
Figure 41 – The Syringe Subscreen on the Heparin Delivery Screen
86
Function
The Heparin on/off key activates the heparin administration program.
It is located on the lower, right side of the control panel. The green
light above the button is illuminated when the pump is running.
When the light is flashing, the operation is interrupted and will resume
when the Treatment Clock is turned on or the Heparin on/off key is
pressed.
If the Heparin on/off key is pressed to turn off the pump, the light will
turn off and the pump will not resume with the Treatment Clock
button.
2008K Hemodialysis Machine Operator's Manual Rev P