Fresenius Medical Care 2008K Operator's Manual page 155

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Chapter 6—Alarms and Troubleshooting
Message
Purpose of Message
BPM: Pulse Amp Unif
The amplitude of the pressure pulses
is inconsistent with an accurate blood
pressure profile.
BPM: Pulse High
The latest pulse reading is above the
Upper Pulse alarm limit.
BPM: Pulse Low
The latest pulse reading is below the
Lower Pulse alarm limit.
BPM: Pump On > 30 sec
The pump that inflates the cuff has
been running longer than 30 seconds
for an adult patient or longer than 10
seconds for pediatric patient.
BPM: Systolic High
The systolic blood pressure reading is
above the set Upper Systolic alarm
limit.
BPM: Systolic Low
The systolic blood pressure reading is
below the set Lower Systolic alarm
limit.
BPM: Weak Pulse
The pulse pressure is too weak to
register an accurate measurement.
BPM: Zero Pressure
No pressure is detected by the blood
pressure module.
BTM test underway
The OLC test may not be started
when a BTM recirculation test is
underway.
BVM Failed
The BVM module has failed
155
Type
Action Required
Blood
Check pressure cuff for proper fit and alignment. Observe the patient
Pressure
for physiologic changes. Treat as patient as symptoms warrant.
Alarm
Blood
Observe the patient for physiologic changes. Treat patient as
Pressure
symptoms warrant.
Alarm
Blood
Observe the patient for physiologic changes. Treat patient as
Pressure
symptoms warrant.
Alarm
Blood
Cuff is not inflating. Check for loose tubing connections or a leak in
Pressure
the cuff.
Alarm
Blood
Observe the patient for physiologic changes. Treat patient as
Pressure
symptoms warrant.
Alarm
Blood
Observe the patient for physiologic changes. Treat patient as
Pressure
symptoms warrant.
Alarm
Blood
Check the cuff for proper fit and inflation. Observe the patient for
Pressure
physiologic changes. Treat patient as symptoms warrant.
Alarm
Blood
Check for a loose connection in the inflation system. Correct as
Pressure
necessary. If no leak found, turn the power off, then back on. If
Alarm
problem persists, alert a qualified service technician.
Advisory
Wait for the BTM recirculation test to complete before beginning the
OLC test.
Alarm
Press RESET to clear the message. BVM will no longer pass
information to the 2008K monitor. Turn the power off and back on. If
the alarm is not cleared, refer to a qualified service technician.
2008K Hemodialysis Machine Operator's Manual Rev P