Fresenius Medical Care 2008K Operator's Manual page 229

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Differences between the 2008K and 2008K@home
Hemodialysis Machines
Figure 74 – 2008K and 2008K@home Control Panels
Table 35 – Control Panel Differences
Item
Essential Control
Panel Keys
Miscellaneous
Control Panel Keys
Indicator Lights
2008K Hemodialysis Machine Operator's Manual Rev P
Appendix C—The 2008K@home Hemodialysis Machine
2008K
Power, Mute,
Stat/Deflate, UF on/off
RESET
Override
Prime, Heparin on/off,
Dialysate Flow on/off,
SVS on/off
New TX
Air Detector, Blood Leak
Detector, Minor, Bypass
2008K@home
Same as 2008K
Press to reset alarms
Press and hold RESET for two seconds
(not one second as on the 2008K
hemodialysis machine) to temporarily
widen the alarm window for arterial,
venous, and transmembrane (TMP)
pressures
During a blood leak alarm, press and
hold RESET for three seconds to
override the alarm and keep the blood
pump running for three minutes
These functions are handled by buttons
on the Touchscreen
On the Data Entry Keypad, press the
Down Arrow () and CONFIRM at the
same time to prompt for a new treatment
These notifications are handled by the
machine's Status Bar; Bypass is also
visible by watching the flow indicator in
the dialysate supply line
229