Fresenius Medical Care 2008K Operator's Manual page 223

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Cuff Pressure Range
Initial Cuff Inflation
Systolic Determination Range
MAP Determination Range
Diastolic Determination Range
Pulse Rate Determination Range
Cuff Inflation Rate
Determination Time Normal
Overpressure Cut Off
Transducer Drift
Leakage Rate (Max)
Pressure Rate Offset
Systolic
MAP
Diastolic
Pulse
Inflation Pressure
2008K Hemodialysis Machine Operator's Manual Rev P
Performance Limits
Pediatric
5-220 mmHg
125 mmHg or adjusted by
host
30-200 mmHg
25-140 mmHg
10-180 mmHg
40-240 BPM
5 seconds
Approx. 20 seconds
220 mmHg
Auto Zeroing
3 mmHg/min in 3 minutes
Auto Zeroing
Alarm Preset Values
Internal alarm values preset to provide alarm limits in
the event individual values are not entered
Pediatric/Neonatal
160/80
120/70
100/50
120/50
Auto
Machine Specifications
Adult
10-325 mmHg
180 mmHg or adjusted by
host
60-260 mmHg
35-220 mmHg
30-200 mmHg
40-180 BPM
5 seconds
25-30 seconds
325 mmHg
Auto Zeroing
3 mmHg/min in 3 minutes
Auto Zeroing
Adult
200/80
120/70
110/50
120/50
Auto
223