Fresenius Medical Care 2008K Operator's Manual page 113

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Table 27 – The BTM Data Screen Features
Feature
Note: When the 2008K hemodialysis machine is first turned on, the small display on the
BTM will indicate 1107. This is a normal event and can be cleared by pressing the Up 
(Error) and Down  (Result) keys on the BTM module at the same time.
2008K Hemodialysis Machine Operator's Manual Rev P
Function
This area will display "Recirculation" when a recirculation
test is being done. When there is no control program, it will
read "Monitoring." It will read "Temperature" or "Energy"
when in a temperature or energy control mode.
When performing a recirculation measure, the data box will
display the T set value that the dialysate will reach.
When in a temperature control mode, this area will display
the rate of temperature change prescribed to warm or cool
the patient in °C/h.
When in an energy control mode, this display will indicate
the energy flux to or from the patient in kilojoules per hour
(kJ/h).
This displays the arterial bloodline temperature as reported
by the BTM module.
This displays the venous bloodline temperature as reported
by the BTM module.
The table above the graph will display
up to 3 recirculation values. This graph
shows the arterial temperature in red
and the venous temperature in blue.
During recirculation tests the
temperature will show changes for a
short period of time. The vertical dotted
line indicates the scheduled end of
treatment.
Chapter 4—Monitoring the Treatment
113