Fresenius Medical Care 2008K Operator's Manual page 168

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

168
Hold CONFIRM to Prime
In heparin screen after selecting
prime must press and hold confirm to
prime heparin line.
I2C Read Time Out
Functional to I2C EEPROM
communication problem.
I2C Bus Read Error
I2C Bus Read Too Long
I2C Byte Write Error
In bypass for 8 min
Press RESET to clear the message
INTERRUPT RINSE?
Press CONFIRM to interrupt rinse.
Escape or CONFIRM
Interrupted
The selected Rinse program has
been interrupted.
Invalid Data Entry for
Entry value for [item] is out of range
[item]
Invalid UF Rate
Entry value for goal is out of range.
Invalid UF Time
Entry value for goal is out of range.
Chapter 6—Alarms and Troubleshooting
2008K Hemodialysis Machine Operator's Manual Rev P
Dialog
Press and hold CONFIRM to prime the heparin line.
Message
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Advisory
The machine was in bypass for about eight minutes. This may
extend the time necessary to complete the treatment or rinsing of
germicide. Press RESET to clear the message.
Warning! If rinsing germicide from the machine or
dialyzer when this occurs, additional time will be
necessary to fully rinse the germicide from the dialyzer.
Always check for residual germicide using the
appropriate approved residual test method.
Dialog
Press CONFIRM to accept or press escape to cancel.
Message
Warning
Re-insert the dialysate connectors into the proper rinse ports. To
continue the Rinse or other program, press CONFIRM, then reselect
the desired program.
Dialog
Set appropriate value for [item]
Message
Dialog
Readjust rate
Message
Dialog
Readjust time
Message