Chapter 4—Monitoring the Treatment
The following is information about each of the information lines in the treatment summary
display.
Table 24 – The Treatment Summary Information
Display
Pt ID
Start Time
End Time
Actual Tx
time
RTD
Hep. Infused
Begin BP
End BP
Diastolic
Range
Systolic
Range
MAP Range
Pulse Range
UF Goal
UF Removed
Avg UF Rate
UF Time
UF Profile
102
Description
This is the patient ID. Currently this is only used by the FDS08
system.
This is the clock time when the Tx Clock button is touched (24 hour
clock).
If the treatment is still underway, this is the projected time for the end
of treatment, based on the current time and RTD. Otherwise it is the
clock time when the treatment actually ended, based on the Tx clock
(24 hour clock).
This is the total treatment time, even if the treatment continued after
RTD counted down to zero (minutes).
The current Remaining Time of Dialysis (minutes).
This is the amount of heparin infused to the patient at this point in time
(ml).
Displays the first Diastolic, Systolic (mmHg) and pulse reading
(beats/min). If any of the readings is out of the alarm range, the entire
line is shown in red.
Displays the last Diastolic, Systolic and pulse reading. If any of the
readings is out of the alarm range, the entire line is shown in red.
Displays the highest and lowest Diastolic pressure reading during the
treatment. If any of the readings is out of the alarm range, the entire
line is shown in red.
Displays the highest and lowest Systolic pressure reading during the
treatment. If any of the readings is out of the alarm range, the entire
line is shown in red.
Displays the highest and lowest Mean Arterial Pressure (MAP) during
the treatment. If any of the readings is out of the alarm range, the
entire line is shown in red (mmHg).
Displays the highest and lowest pulse rate during the treatment. If any
of the readings is out of the alarm range, the entire line is shown in
red.
This is the UF goal selected for the treatment (ml).
This is the UF removed at this point in the treatment (ml).
This is the average UF rate at this point in the treatment (ml).
This is the UF time selected for the treatment (min).
This is the UF profile # selected for the treatment
2008K Hemodialysis Machine Operator's Manual Rev P