158
Conductivity out of range
When attempting to start an OLC test,
certain conditions are necessary,
including the condition that
conductivity for both the inlet and
outlet sensors be in range
CONFIRM Concentrate
This message will be displayed if the
user needs to confirm the concentrate
selected for use
Connector(s) Out Of Port
An action has been initiated that
requires the Acid/Bicarbonate
Connectors to be in their rinse ports
Cooling Down
The machine is cooling down from a
heat disinfect.
** Cover is Open **
The dialysate shunt door is open.
Cuff Pressure = XXX
This is displayed during the blood
pressure measurement. The cuff
pressure is XXX mmHg.
Dial Valve Failure 1
Electronic self-test failure.
Dial Valve Failure 2
Electronic self-test failure.
Chapter 6—Alarms and Troubleshooting
2008K Hemodialysis Machine Operator's Manual Rev P
Advisory
Wait until the conductivity is stable and start the OLC test again. If
the message repeats, do not use OLC until the conductivity sensors
have been recalibrated.
Advisory
Press CONFIRM or change the concentrate selection and then press
CONFIRM.
Warning
Insert the concentrate connectors into their proper rinse ports.
Advisory
Advisory message only. No action is required.
Advisory
To proceed with the selected operation, close the shunt door.
Dialog
No action is necessary.
Message
Alarm
Turn machine power off and back on. If alarm does not clear, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If alarm does not clear, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.