218
Current national (U.S.) Standards for the Quality of Water:
ANSI/AAMI 13959:2014, Water for hemodialysis and related
therapies
ANSI/AAMI 26722:2014, Water treatment equipment for
hemodialysis applications and related therapies
Other related standards include:
ANSI/AAMI RD62:2006, Water treatment equipment for hemodialysis
applications and related therapies
3 feet maximum height. Must comply with local codes and must maintain
a free fall air gap between drain hose and building drain.
3 meters (approximately 10 feet) maximum drain hose length
Temperature 37 °C. Flow rate 620 ml/min. Time between 10 and 60
minutes (internally selectable)
Current national (U.S.) Standards for the Quality of Dialysis Fluid:
ANSI/AAMI 11663:2014, Quality of dialysis fluid for hemodialysis and
related therapies
ANSI/AAMI 23500:2014, Guidance for the preparation and quality
management of fluids for hemodialysis and related therapies
Other related standards include:
ANSI/AAMI RD52:2004, Dialysate for hemodialysis
Dialysate on/off key
Accuracy: ± 5%
Sequential (0)/100/200/300/400/500/600/700/800
ml/ min., selectable in the Dialysis screen; Additionally: 1.5X or 2.0X
dialysate flow rate based on the Blood Pump rate (Qb):
Qb w/1.5X selected
0 – 165
166 – 215
216 – 315
315 and below
316 – 415
416 – 480
481 and above
Note: All flow rates are approximate. Dialysate flow will not adjust unless
the blood pump is adjusted at least 15 – 20 ml/min.
† (if Auto Flow Minimum 300 Qd is set in Service Mode)
‡ (if Auto Flow Minimum 500 Qd is set in Service Mode)
From Drain line, intermittent collection using a 3 Liter PD drain bag as a
collection device with a Safe-Lock connector (optional).
Current national (U.S.) Standards for the Quality of Concentrates: ANSI/
AAMI 13958:2014, Concentrates for hemodialysis and related therapies
Max suction height 3 feet; Max supplied pressure 2 psi
2008K Hemodialysis Machine Operator's Manual Rev P
Qb w/2.0X selected
0 – 150
151 – 215
216 – 265
265 and below
266 – 315
316 – 365
366 and above
Qd
300
400
500
500
600
700
800