Fresenius Medical Care 2008K Operator's Manual page 215

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Other Options
There are other options on these screens that may not be specifically described here.
Generally, these are setup options that are based upon the presence or absence of other
modules or hardware.
Options which require a "key" for activation
Certain options are extra cost or limited in availability and must be activated by use of a
special code stored on the functional board. Examples of this sort of feature include OLC.
Upon purchase of such a feature, use the following procedure to activate it on the machine.
Caution: This procedure should be performed by a qualified individual. Be sure to follow
electrostatic discharge (ESD) procedures when removing the board and replacing the
EEPROM chips on the board.
To transfer the option, the EEPROM "hardware key" is placed in IC20 on the
functional board in place of the calibration EEPROM. Save the calibration EEPROM
as it must be returned to the same machine. Be sure to match the notch on the
EEPROM with the socket.
Return the functional board to the machine and turn on the power.
Upon power up the functional board reads the contents of the memory from the
EEPROM key.
The message "New feature loaded. Power off. Replace eeprom." is displayed.
Turn off power, replace the EEPROM "hardware key" with the calibration EEPROM,
and turn the power back on.
Note: The key may be used only once. If the key is reused, the machine locks up and
displays the message "Eeprom already used. Power off. Replace eeprom".
2008K Hemodialysis Machine Operator's Manual Rev P
Appendix B
215