Set Venous Limits
This is a message to adjust the
venous limits if necessary
Short Power Down
The machine was switched off for 1 –
2 minutes and turned back on. Setup
values have not been set to default
values.
Single Needle On!
An action has been initiated that
requires the Single Needle option to
be off.
SN BP +5 V Error
+ 5 volts is outside the allowable
range
SN BP +12 V Error
+ 12 volts is outside the allowable
range
SN BP -12 V Error
- 12 volts is outside the allowable
range
SN BP +24 V Error
+ 24 volts is outside the allowable
range
SN BP Button Alarm
Key stuck or held in too long
SN BP Comm. Timeout
Time out when receiving Intel-Hex-
line or overflowed received buffer
SN BP Del. Rate Alarm
Actual speed-read back analog
voltage at X348/10 is out of limits
SN BP Direction Alarm
Pump is turning in the wrong direction
SN BP EEPROM Error
EEPROM error
SN BP EPROM Error
EPROM CRC error
SN BP Erasing Error
Error erasing Flash ROM while in
Service Mode
SN BP Flash Error
Error copying data into Flash ROM
while in Service Mode
SN BP Level Up Alarm
Pressure increase when the Level Up
key is pressed
Chapter 6—Alarms and Troubleshooting
2008K Hemodialysis Machine Operator's Manual Rev P
Advisory
Press and hold RESET for 1 second to adjust the limits. Changes in
venous pressure during the treatment should be investigated. See
"Venous Pressure Alarm"
Advisory
Verify that the dialysis parameters are as desired.
Warning
To proceed with the selected operation, de-select the Single Needle
option.
Alarm
See message E.10
Alarm
See message E.07
Alarm
See message E.09
Alarm
See message E.08
Alarm
See message A.16
Alarm
See message A.27
Alarm
See message A.21
Alarm
See message A.13
Alarm
See message E.05
Alarm
See message E.01
Alarm
See message E.98
Alarm
See message E.97
Alarm
See message A.25