Chapter 6—Alarms and Troubleshooting
Message
Purpose of Message
UF Pump Alarm
UF pump is not connected or is not
pulsing properly.
UF Rate Error
A calculation error has been detected
UF Removed cleared
This temporary message is displayed
when the Tx clock is turned on the
first time after the New TX key was
pressed. The UF removed has been
set to 0.
UF Removed not cleared
This temporary message is displayed
when the Tx clock is turned on other
than the first time after the New TX
key was pressed. The UF removed
has been not been set to 0.
Upper Dia. has been set
The operator has attempted to set the
to Min [Max]
upper diastolic pressure limit higher or
lower than allowed. The machine has
set the limit to the highest or lowest
value allowed.
Upper Pulse has been set
The operator has attempted to set the
to Min [Max]
upper pulse rate limit higher or lower
than allowed.
Upper Sys. has been set
The operator has attempted to set the
to Min [Max]
upper systolic pressure limit higher or
lower than allowed.
Valve 43 Failure
Valve 43 has remained open too long
183
Type
Action Required
Alarm
Press RESET to reset the alarm. If unable to clear the alarm, return
the blood to the patient if alarm occurs during treatment. Take
machine out of service and alert a qualified service technician.
Alarm
Reset the UF parameters
Advisory
No action necessary.
Advisory
No action necessary.
Dialog
Verify that the limit setting is acceptable
Message
Dialog
The machine has set the limit to the highest or lowest value allowed.
Message
Verify that the limit setting is acceptable
Dialog
The machine has set the limit to the highest or lowest value allowed.
Message
Verify that the limit setting is acceptable
Alarm
Turn machine power off and back on. Just before beginning dialysis,
verify that the dialysate flow can be turned off and back on. Do not
initiate or continue dialysis if this cannot be done.
2008K Hemodialysis Machine Operator's Manual Rev P