Chapter 6—Alarms and Troubleshooting
Message
Purpose of Message
A.25 (Arterial or SN
Pressure increase when the Level Up
Blood Pump Message)
key is pressed
A.26 (Arterial Blood
Pressure was adjusted too much in
Pump Message)
calibration mode
A.27 (Arterial Blood
Time out when receiving Intel-Hex-
Pump Message)
line or overflowed received buffer
A.28 (Arterial Blood
Error in received Intel-Hex-line
Pump Message)
A.29 (Arterial Blood
Pump rotor turning when it should not
Pump Message)
be
Access Flow Complete
This message is an advisory
message that the Access Flow test is
complete
Access Flow Running
This message is an advisory
message that the Access Flow test
process is continuing.
Access Flow Test
This message is an advisory
Scheduled
message that the Access Flow test
process is continuing.
Acetate Selected!
Acetate concentrate has been
selected and the blue bicarbonate
wand/connector is out of its port.
145
Type
Action Required
Alarm
Press the RESET key to clear. Possibility that the level adjust pump
is connected backward so that the level is lowered instead of raised.
Verify that the level in the arterial chamber rises when the adjust key
is pressed. If it does not, return the blood to the patient if alarm
occurs during a treatment. Take the machine out of service and alert
a qualified service technician.
Alarm
Press the RESET key if this message occurs in dialysis mode. If this
alarm occurs during a treatment, return the blood to the patient. Take
the blood pump module out of service and alert a qualified service
technician.
Alarm
Press the RESET key if this message occurs in dialysis mode. If this
alarm occurs during a treatment, return the blood to the patient. Take
the blood pump module out of service and alert a qualified service
technician.
Alarm
Press the RESET key if this message occurs in dialysis mode. If this
alarm occurs during a treatment, return the blood to the patient. Take
the blood pump module out of service and alert a qualified service
technician.
Alarm
Press the RESET key to clear. If problem persists, manually return
the blood to the patient (see page 118 for instructions). Take blood
pump module out of service and alert a qualified service technician.
Warning
Press CONFIRM to clear the message
Advisory
No action is necessary
Advisory
No action is necessary
Warning
Connect blue (bicarbonate) wand/connector into the blue rinse port.
Be sure the concentrate selection is correct.
2008K Hemodialysis Machine Operator's Manual Rev P