Chapter 6—Alarms and Troubleshooting
Message
Purpose of Message
# of Tests has been set to
The operator has attempted to set the
min
number of Online Clearance (OLC)
tests lower than allowed.
5 Minutes Flow Off
Dialysate flow has been off for five
minutes.
*** 5V HIGH ***
Electronic self-test, power supply
limits exceeded.
*** 5V LOW ***
Electronic self-test, power supply
limits exceeded.
10 Fill Pgm in 1 hr
Ten fill programs have occurred
during a one-hour period.
12V POWER FAIL
Electronic self-test, power supply
limits exceeded.
*** 24 V HIGH
Electronic self-test, power supply
limits exceeded.
*** 24V LOW
Electronic self-test, power supply
limits exceeded.
143
Type
Action Required
Dialog
The machine has set the number of tests to the lowest value allowed.
Message
Verify that the number of OLC tests is acceptable. See page 104 for
more information.
Warning
1) Press RESET to silence the alarm.
2) If you intend for the flow to be off, set the Dialysate Flow button
to Seq(uential). Otherwise, turn on the dialysate flow with the
On/Off key.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Warning
1) Check the dialyzer supply and return lines, especially around the
connectors and dialysate filter in the dialyzer return line, for air
entering the system and correct the problem.
2) Press RESET to clear the alarm. If unable to reset the alarm,
return the blood to the patient, take the machine out of service
and replace the machine with another machine. Alert a qualified
service technician.
Note: Using a conventional dialyzer at a high UF rate can
cause frequent Fill programs because of a high TMP.
Lowering the UF rate by decreasing the UF Goal may
solve the problem. Notify a physician if the UF goal has
changed.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
2008K Hemodialysis Machine Operator's Manual Rev P