Fresenius Medical Care 2008K Operator's Manual page 174

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

174
Online Clearance Test
Machine is running an OLC
measurement
Online PHT Failed
The online Pressure Holding Test has
failed.
Online PHT Too Long
Electronic self-test failure.
Patient Alarm
External alarm
PHT is running
The PHT test must complete before
the OLC test is allowed to run.
Plug in Venous Pump
Single Needle option was initiated but
the single-needle, blood pump is not
plugged into the machine port.
Power Failure Recovery
The machine is powering up after a
power failure. Parameters have been
recovered
Press CONFIRM to exit
This message is a prompt for the
operator to press the CONFIRM key
to exit the Rinse program.
Press CONFIRM to Load
This message is a prompt for the
operator to press the CONFIRM key
to load the heparin syringe.
Press CONFIRM to Start
This message is a prompt for the
operator to press the CONFIRM key
to start the program.
Press ESCAPE to cancel
This message is a prompt for the
rinse
operator to press the Escape key to
cancel the Rinse program.
Chapter 6—Alarms and Troubleshooting
2008K Hemodialysis Machine Operator's Manual Rev P
Advisory
Advisory only no action required
Warning
Reset the alarm. Check the machine for liquid leaks. If the failure
message is repeated on the next test (12 minutes between tests),
return the blood to the patient if alarm occurs during treatment. Take
the machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Alarm
Clear external alarm. If problem persists, return the blood to the
patient if alarm occurs during treatment. Take the machine out of
service and alert a qualified service technician.
Advisory
Wait 15 seconds and start the OLC test again.
Warning
To proceed with the selected operation, install the single-needle
blood pump into the machine.
Opening
Verify that all treatment settings are correct before resuming dialysis.
Screen
Message
Advisory
To proceed with the selected operation, press CONFIRM.
Dialog
To proceed with the selected operation, press CONFIRM.
Message
Advisory
To proceed with the selected operation, press CONFIRM.
Advisory
To proceed with the selected operation, press Escape then press
CONFIRM.