Fresenius Medical Care 2008K Operator's Manual page 176

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

176
Recirculating Done
A prompt to the operator that the
recirculation process is done.
Recirculating Stopped
Recirculation has been stopped
because blood is sensed or the
dialyzer supply and return lines are on
shunt.
Release CONFIRM to
This message is a prompt for the
stop
operator to release the CONFIRM key
to stop priming the heparin line.
Rel. Blood Volume Low
The BVM module has reported a
relative blood volume below the lower
limit.
RESET to adjust TMP
The TMP exceeded the hard alarm
limits. The operator is given the
option to relieve the pressure to bring
the TMP within limits.
Reset Treatment?
The New TX key has been pressed
CONFIRM or Escape
Resetting, Try Again
Blood pressure module resetting
Retry > Press = XXX
The cuff pressure is too low to
measure the blood pressure. The cuff
pressure is XXX mmHg.
Reverse bloodlines
This message is a prompt for the
operator reverse the bloodlines for the
Access Flow test
Rinse Cond High
The Reverse Osmosis (RO) water
inlet conductivity is too high.
Chapter 6—Alarms and Troubleshooting
Advisory
Warning
Message
Warning
Warning! Adjusting the
TMP repeatedly will
decrease the UF removed
from the patient.
Advisory
Warning
Message
Warning
Opening
Message
2008K Hemodialysis Machine Operator's Manual Rev P
Press RESET to clear the advisory message.
Inspect the configuration of the dialyzer supply and return lines and
extracorporeal blood circuit. Correct any irregularities. If the message
is not cleared, take the machine out of service and alert a qualified
service technician.
Dialog
Release the CONFIRM key.
Alarm
Press RESET to clear the message. Evaluate the fluid status of the
patient.
Press the RESET key to reset the TMP alarm limits. Press and hold
the Override key to re-center the limits.
Warning! Rising TMP may indicate a leak in the balancing
system and should be investigated.
To reset treatment parameters for a new treatment, press the
CONFIRM key. To cancel, press the Escape key.
Wait until blood pressure module completes resetting and retry blood
pressure reading.
Dialog
No action necessary
To proceed with the Access Flow test, reverse the bloodlines and
press CONFIRM.
Press the RESET key to clear the message. Perform a Rinse cycle. If
Screen
alarm is not cleared, take the machine out of service and alert a
qualified service technician.