Chapter 3—Setting Treatment Parameters
Setting the Acid/Bicarbonate Alert
The Acid/Bicarbonate Alert option sounds an alarm when the fluid level in either of the
concentrate jugs has been drained to 20 percent of its original amount. In addition to the
alarm, a warning message such as LOW ACID WARNING , LOW BIC WARNING , or LOW
ACETATE WARNING will appear in the status box. This alert aids the operator in
maintaining adequate amounts of concentrate in the containers during treatment. Be sure to
set the new volume in this screen whenever the concentrate containers are refilled.
Note: If the facility is
on a centralized acid
or bicarbonate supply
system, enter zero to
deactivate the alarm
for that concentrate.
Figure 31 – Setting Acid & Bicarbonate Alerts
68
2) Touch Acid. Using the data entry keypad, enter the
actual amount, in liters, of concentrate in the acid jug.
3) Touch Bicarbonate. Using the data entry
keypad, enter the actual amount, in liters, of
concentrate in the bicarbonate jug.
1) From the "Dialysate"
screen, touch
Acid/Bicarb Alert. The
"Acid/Bicarb Alert"
subscreen appears.
4) Touch Alert until an X
appears in the On checkbox.
5) Press CONFIRM to
save the input data.
6) Touch Dialysate to close the "Acid/Bicarb
Alert" subscreen. The volume of each
concentrate will count down as it is used.
2008K Hemodialysis Machine Operator's Manual Rev P