Fresenius Medical Care 2008K Operator's Manual page 164

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

164
Fail * Ven High Soft
Venous Pressure test has failed.
Fail * Ven Low Soft
Fail * Ven High Hard
Fail * Ven Low Hard
** Failed Sending Data to
Functional to Actuator board
Actuator Board **
communication problem during
startup.
Fill for Diasafe Test
This message indicates the status of
the Diasafe test
Fill Program Alarm
A Fill program has occurred for one
minute while blood is sensed.
Filling Program
A Fill program is in progress.
Flow Error
General Flow Alarm
Chapter 6—Alarms and Troubleshooting
2008K Hemodialysis Machine Operator's Manual Rev P
Test
Rerun test, if failure repeats, remove machine from service and alert
Message
a qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Advisory
Advisory only. No action is required.
Alarm
Inspect for air in the system.
Correct as required.
If the warning is not cleared, return the blood to the patient if alarm
occurs during treatment. Take the machine out of service and alert a
qualified service technician.
Advisory
Advisory only. No action is required.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly:
1) Check the water supply flow to the machine.
2) Check that the Dialysate Flow is on.
3) Check the dialyzer supply and return lines for kinks.
4) Set Dialysate Flow in the "Home" screen to 500 ml/min and verify
that the flow from the drain line is 500 ml/min  50 ml/min.
5) Turn the power off and on. If warning is not cleared, return the
blood to the patient if alarm occurs during treatment.
6) Take the machine out of service and alert a qualified service
technician.