To add a Citrasate® concentrate that is not in the catalog (Functional board software
version 3.36 or later):
On the "Add New Concentrate" screen, set the Citrasate® option to Yes and press
CONFIRM. The Acetic Acid button will change to a Citrate meter-box. Modify all of the
constituents to match the concentrate. Select the Conc Name button and then input the
desired alphanumeric name. Use the number keys on the front panel for the numeric values
and the letters on the screen for the alphabetic values. Press CONFIRM to confirm your
changes. After all of the constituents are correctly entered and the name of the concentrate
has been chosen, select the Enter Conc screen-button to verify that the concentrate was
entered into the list.
Note: If an acetate concentrate is selected, the Citrasate® option is not available.
In Dialysis Mode, the dialysate composition list will now show a meter box for 'Citrate'.
Figure 70 – Dialysate Composition with Citrate on the "Dialysate" screen
2008K Hemodialysis Machine Operator's Manual Rev P
Citrate
meter
box
Appendix B
207