Chapter 6—Alarms and Troubleshooting
Message
Purpose of Message
Flow Inlet Error
Float Switch
Flow is Off
Dialysate flow is off.
Flow is still On
Dialysate flow is on.
Flow Rate not Set
If the dialysate flow is turned on while
the touch screen flow rate selection is
still "SEQ", this reminder is displayed
Flow Recirc Error 1
Dialysate flow problem.
165
Type
Action Required
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly:
1) Check the water supply flow to the machine.
2) Check that the Dialysate Flow is on.
3) Set Dialysate Flow in the "Home" screen to 500 ml/min and verify
that the flow from the drain line is 500 ml/min  50 ml/min.
4) Turn the power off and on. If warning is not cleared, return the
blood to the patient if alarm occurs during treatment.
5) Take the machine out of service and alert a qualified service
technician.
Warning
An action has been initiated that requires the dialysate flow to be on.
To proceed with the selected operation, turn the dialysate flow on.
Warning
An action has been initiated that requires the dialysate flow to be off.
To proceed with the selected operation, first turn the dialysate flow
off.
Advisory
Set the dialysate flow rate to the desired value.
Warning
A single occurrence is not a problem if the machine automatically
resets. If the problem persists for longer than one minute or occurs
repeatedly:
1) Check the water supply flow to the machine.
2) Check that the Dialysate Flow is on.
3) Set Dialysate Flow in the "Home" screen to 500 ml/min and verify
that the flow from the drain line is 500 ml/min  50 ml/min.
4) Turn the power off and on. If warning is not cleared, return the
blood to the patient if alarm occurs during treatment.
5) Take the machine out of service and alert a qualified service
technician.