Chapter 6—Alarms and Troubleshooting
Message
Purpose of Message
Press ESCAPE To Stop
This message is a prompt for the
[Item]
operator to press the Escape key to
stop loading the heparin syringe or
the Rinse program.
Pressure Test Failed
The pressure test section (PHT) of
the automated Test Sequence has
failed.
Priming
The operator has pressed the Prime
key and initiated the priming function.
Put Connectors in Port
The connectors must be in the
machine ports in order to start a
Rinse program
Put Lines On Shunt
An action has been initiated that
requires the dialyzer supply and
return lines to be on the shunt.
Put Red Con in Chemical
This is a cleaning/disinfectant
program prompt to the operator.
RAM Battery Failure
Electronic self-test failure.
RAM Code Corrupted 1
Electronic self-test failure.
RAM Code Corrupted 2
Repeated electronic self-test failure.
Recirc Interrupted
The Recirculate program has been
interrupted by an alarm condition.
Recirculating
Recirculation is in progress.
175
Type
Action Required
Dialog
To proceed with the selected operation, press Escape then press
Message
CONFIRM.
Alarm
Reset the alarm and repeat the test. If the failure message is
repeated on retest, take the machine out of service and alert a
qualified service technician.
Advisory
Advisory only. No action is required.
Advisory
Connect the red (acid/acetate) and/or blue (bicarbonate) connectors
to the appropriate rinse ports.
Warning
To proceed with the selected operation, place dialyzer supply and
return lines on the shunt.
Advisory
Remove the red connector from the machine and place it into the
wand in the yellow chemical/disinfectant bottle.
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, return
the blood to the patient if alarm occurs during treatment. Take the
machine out of service and alert a qualified service technician.
Warning
1) Inspect the blood pump condition. Correct if required.
2) Reset the alarm and turn UF back on, if applicable.
3) If the alarm does not clear, take the machine out of service and
alert a qualified service technician.
Advisory
Advisory only. No action is required.
2008K Hemodialysis Machine Operator's Manual Rev P